Technical data
734 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
33 Obsolete and Discontinued Commands
:MEASure:VDELta
 (see page 790)
Query Syntax
:MEASure:VDELta?
The :MEASure:VDELta? query returns the voltage difference between 
vertical marker 1 (Y1 cursor) and vertical marker 2 (Y2 cursor). No 
measurement is made when the :MEASure:VDELta? query is received by 
the oscilloscope. The delta value that is returned is the current value. This 
is the same value as the front- panel cursors delta Y value.
VDELta = value at marker 2 - value at marker 1
Return Format
<value><NL>
<value> ::= delta V value in NR1 format
See Also • "Introduction to :MARKer Commands" on page 318
• "Introduction to :MEASure Commands" on page 340
• ":MARKer:Y1Position"  on page 327
• ":MARKer:Y2Position"  on page 328
• ":MARKer:YDELta"  on page 329
• ":MEASure:TDELta"  on page 727
• ":MEASure:TSTArt"  on page 729
NOTE
The :MEASure:VDELta command is an obsolete command provided for compatibility to 
previous oscilloscopes. Use the :MARKer:YDELta command (see page 329) instead.