35 Status Reporting
• Table 56
• Table 51
The status registers picture above shows how the different status reporting 
data structures work together. To make it possible for any of the Standard 
Event Status Register bits to generate a summary bit, the bits must be 
enabled. These bits are enabled by using the *ESE common command to 
set the corresponding bit in the Standard Event Status Enable Register.
To generate a service request (SRQ) interrupt to an external controller, at 
least one bit in the Status Byte Register must be enabled. These bits are 
enabled by using the *SRE common command to set the corresponding bit 
in the Service Request Enable Register. These enabled bits can then set 
RQS and MSS (bit 6) in the Status Byte Register.