Technical data
770 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
35 Status Reporting
(Standard) Event Status Enable Register (ESE)
To allow any of the (Standard) Event Status Register (ESR) bits to 
generate a summary bit, you must first enable that bit. Enable the bit by 
using the *ESE (Event Status Enable) common command to set the 
corresponding bit in the (Standard) Event Status Enable Register (ESE).
Set bits are read with the *ESE? query.
Example Suppose your application requires an interrupt whenever any type of error 
occurs. The error related bits in the (Standard) Event Status Register are 
bits 2 through 5 (hexadecimal value 0x3C). Therefore, you can enable any 
of these bits to generate the summary bit by sending:
myScope.WriteString "*ESE " + CStr(CInt("&H3C"))
Whenever an error occurs, it sets one of these bits in the (Standard) Event 
Status Register. Because all the error related bits are enabled, a summary 
bit is generated to set bit 5 (ESB) in the Status Byte Register.
If bit 5 (ESB) in the Status Byte Register is enabled (via the *SRE 
command), an SRQ service request interrupt is sent to the controller PC.
NOTE
Disabled (Standard) Event Status Register bits respond but do not generate a summary 
bit. (Standard) Event Status Register bits that are not enabled still respond to their 
corresponding conditions (that is, they are set if the corresponding event occurs). However, 
because they are not enabled, they do not generate a summary bit to the Status Byte 
Register.