Technical data
774 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
35 Status Reporting
Arm Event Register (AER)
This register sets bit 5 (Wait Trig bit) in the Operation Status Register and 
the OPER bit (bit 7) in the Status Byte Register when the instrument 
becomes armed.
The ARM event register stays set until it is cleared by reading the register 
with the AER? query or using the *CLS command. If your application 
needs to detect multiple triggers, the ARM event register must be cleared 
after each one.
If you are using the Service Request to interrupt a program or controller 
operation when the trigger bit is set, then you must clear the event 
register after each time it has been set.