Specifications
24
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics 
 * Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 ohms. 
 ** Sinc, Cardiac and Gaussian Pulse: ±1.25 V into Hi-Z; +- 625 mV into 50 ohms 
*** Full resolution is not available at output due to internal attenuator stepping.
WaveGen – Built-in function/arbitrary waveform generator (specifications are typical) (Continued)
Frequency •  Sine wave and ramp accuracy:
 ◦ 130 ppm (frequency < 10 kHz)
 ◦ 50 ppm (frequency > 10 kHz)
•  Square wave and pulse accuracy:
 ◦ [50+frequency/200] ppm (frequency < 25 kHz)
 ◦ 50 ppm (frequency ≥ 25 kHz)
•  Resolution: 0.1 Hz or 4 digits, whichever is larger 
Amplitude •  Range:
 ◦ 20 mVpp to 5 Vpp into Hi-Z**
 ◦ 10 mVpp to 2.5 Vpp into 50 ohms**
•  Resolution: 100 µV or 3 digits, whichever is higher
•  Accuracy: 2% (frequency = 1 kHz) 
DC offset •  Range:
 ◦ ±2.5 V into Hi-Z**
 ◦ ±1.25 V into 50 ohms**
•  Resolution: 100 µV or 3 digits, whichever is higher
•  Accuracy (waveform modes): ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
•  Accuracy (DC mode): ± 1.5% of offset setting ± 3 mV
Trigger output Trigger output available on Trig out BNC
Main Output • Impedance : 50 ohms typical
• Isolation: not available, main output BNC is grounded
• Protection: overload automatically disables output
Digital voltmeter (specifications are typical)
Functions ACrms, DC, DCrms, Frequency
Resolution ACV/DCV: 3 digits Frequency: 5.5 digits
Measuring rate 100 times/ second
Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements.
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds
.
Connectivity
Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol. 
Two USB 2.0 hi-speed host ports, front and rear panel 
Supports memory devices, printers and keyboards
Optional ports GPIB, LAN (10/100Base-T), WVGA video out
Trigger out BNC connector on the rear panel. Supported modes: triggers, mask, and waveform generator 
sync pulse