Programming instructions
36 
Programming Reference
Programming the Instrument Trigger Modes
Use the :ARM subsystem to select the overall triggering mode of the 
instrument (
CONTINUOUS
, 
TRIGGERED
, 
GATED
, 
EXT WIDTH
), and the 
:TRIGger
 subsystem to select the pulse period source, triggering and 
number of pulse periods per 
:ARM
 event (
BURST
 or 
PATTERN
 length).
Continuous
Set Continuous mode by arming the instrument from its internal 
oscillator:
:ARM:SOURce IMMediate Arm from internal oscillator.
Triggered
Set Triggered mode by arming the instrument on edges from the EXT 
INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT
:ARM:SENSe EDGE Arm on edge
:ARM:SLOPe POSitive Arm on positive edge
:ARM:LEVel 1V Set EXT INPUT threshold
You can also arm the instrument from the PLL and set the frequency (or 
period) of the PLL to the required triggering rate:
:ARM:SOURce INTernal2  Arm from PLL
:ARM:SENSe EDGE  Arm on edge
:ARM:SLOPe POSitive Arm on positive edge
:ARM:FREQuency <value> Set PLL frequency
NOTE The PLL (INTernal2) cannot be used as 
:ARM:SOURce
 (triggering rate) if 
it is already being used as 
:TRIGger:SOURce
 (pulse period source).
Gated
Set Gated mode by arming the instrument on levels from the EXT INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT
:ARM:SENSe LEVel  Arm on signal level
:ARM:SLOPe POSitive Arm on positive level