Programming instructions
42 
Programming Reference
SCPI Instrument Command List
Command :ARM:SENS
Long
:ARM[:SEQuence[1]|STARt][:LAYer]:SENSe
Form Set & Query
Parameter
EDGE | LEVel
*RST value
EDGE
Description Use this command to select Triggered or Gated mode by choosing 
whether the instrument arms on the edge(s) or level of the arming signal. 
When sensing edges, the instrument triggers when the arming signal 
crosses the selected threshold level (
:ARM:LEV
) in the selected direction 
(
:ARM:SLOP
). This corresponds to the Triggered mode selected on the 
Mode/Trigger screen when using the front panel.
When sensing levels, the instrument triggers as long as the arming signal 
is above (
:ARM:SLOP POS
), or below (
:ARM:SLOP NEG
) the selected 
threshold level (
:ARM:LEV
). This corresponds to the Gated mode 
selected on the M
ODE
/T
RIGGER
SCREEN
 when using the front panel.
Command :ARM:SLOP
Long
:ARM[:SEQuence[1]|STAR7t][:LAYer]:SLOPe
Form Set & Query
Parameter
POSitive | NEGative | EITHer
*RST value
POS
Description Use this command to select the trigger slope for the arming signal when 
triggering on edges. Use 
EITHer
 to trigger on both the positive and 
negative edges of the arming signal. This allows you to trigger at twice 
the frequency of the arming signal.
If you are arming on levels, use this command to select whether the 
instrument triggers during the positive or negative cycle of the arming 
signal.