Programming instructions
50 
Programming Reference
SCPI Instrument Command List
Range coupling Amplitude
Description Use this command to program the offset current of the OUTPUT signal. 
Note that to set the OUTPUT levels in terms of current, you first have to 
execute the 
[:SOURce]:HOLD CURRent
 command to enable the 
:SOURce]:CURRent
 subsystem.
The available current range is limited by the combination of:
 •  Specified Voltage limits
 •  Actual OUTPUT Impedance setting 
:OUTPut:IMPedance
 • Actual Expected Load impedance setting
Example To program the offset current of the output signal:
Command :CURR[1]:HIGH
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Form Set & Query
Parameter Numeric
Parameter suffix
A
 with engineering prefixes.
Value coupling
Range coupling Low level
*RST value +10 mA (50 
Ω
 into 50 
Ω
)
Specified Limits  10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50 
Ω 
into short
): 
–82 mA to 152 mA typical 
Description Use this command to program the high level current of the OUTPUT 
signal. Note that to set the OUTPUT levels in terms of current, you first 
:HOLD CURR
:CURR:OFF 50MA
Enable CURRENT subsystem
Set OUTPUT offset to 50 mA
Amplitude = High – Low
Offset = 
High – Low
2