Programming instructions
 53
Programming Reference
SCPI Instrument Command List
Command :CURR[1]:LIM:LOW
Long
[:SOURce]:CURRent[1]:LIMit:LOW
Form Set & Query
Parameter Numeric
Parameter suffix
A
 with engineering prefixes.
*RST value –10.0 mA
Description Use this command to set/read the low level current limit. If you switch on 
current limiting, the low level current cannot be set below the 
programmed limit.
The current is not limited by the OUTPUT hardware, this is a software 
limit. 
Example To set the low level current limit for the output:
Command :CURR[1]:LIM:STAT
Long
[:SOURce]:CURRent[1]:LIMit:STATe
Form Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
OFF
Description Use this command to switch the output limits on or off. When you switch 
on the output limits, you cannot program the output levels beyond the 
programmed limits, until you switch off the output limits. The limits 
apply whether you program high/low levels or amplitude/offset levels.
NOTE You can switch the limits on and off in both the 
[:SOURce]:CURRent
and the 
[:SOURce]:VOLTage
 subsystems but the current and voltage 
limits are not enabled/disabled independently. The voltage and current 
limits are always enabled/disabled together. 
:HOLD CURR
:CURR:LIM:LOW –50MA
:CURR:LIM:STAT ON
Enable CURRENT subsystem 
Set OUTPUT low level current limit to –50 mA
Switch on OUTPUT limits