Programming instructions
56 
Programming Reference
SCPI Instrument Command List
Command :PHAS[1]
Long
[:SOURce]:PHASe[1][:ADJust]
Form Set & Query
Parameter Numeric
Parameter suffix
DEG
 or 
RAD
. A parameter without a suffix is interpreted as 
RAD
.
Functional 
coupling
Programming the pulse phase also executes 
[:SOURce]:PULSe:HOLD 
PHASe
 so that the pulse phase is held constant when the signal frequency 
is changed.
Value coupling
*RST value 0.0
Specified limits 0 to 360
° 
constrained by delay and period limits.
Description Use this command to set/read the relative phase delay of the output 
signal. This is equivalent to setting an absolute or percentage pulse delay 
with 
[:SOURce]:PULSe:DELay
.
If you want the phase delay to remain constant when the pulse period is 
varied (rather than the absolute pulse delay) use 
[:SOURce]:PULSe:DELay[1]:HOLD PRATio
.
Example To set and hold the phase delay: 
D
elay  = 
 Phase
 360
 Period
×
:PHAS 180 DEG
:PULS:DEL:HOLD PRAT
Set OUTPUT phase to 180
°
Hold OUTPUT phase constant with varying 
period