Programming instructions
70 
Programming Reference
SCPI Instrument Command List
Command :VOLT[1]
Long
[:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]
Form Set & Query
Parameter Numeric
Parameter suffix
V
 with engineering prefixes.
Value coupling
Range coupling With Offset, see page 71
*RST value 1.00 V
Specified limits 100 mVpp to 10.0 Vpp (values are valid from 50 
Ω
 into 50 
Ω
)
Description Use this command to program the amplitude voltage of the output signal. 
Note that to set the output levels in terms of voltage, you first have to 
execute the 
[:SOURce]:HOLD VOLTage 
command to enable the 
[:SOURce]:VOLTage
 subsystem.
The available voltage range is limited by the combination of: 
 •  Specified current limits
 •  Actual output impedance setting 
:OUTPut:IMPedance
 •  Actual expected load impedance setting:
:OUTput:IMPedance:EXTernal
Example To set the amplitude voltage:
High = Offset + 
Amplitude
2
Low = Offset – 
2
Amplitude
:HOLD VOLT
:VOLT 5V
Enable VOLTAGE subsystem
Set OUTPUT amplitude to 5 V