Programming instructions
94 
Specifications
Agilent 81101A Specifications
Width
Can be entered as absolute width, duty cycle or trailing edge delay.
Duty cycle values from 0.1% to 95% can be entered directly. For values 
>95 % press SHIFT and use the Modify knob. Note that pulses may be 
deteriorated or skipped due to the inaccuracy of period and width. Hence 
for large values, it is better to select complement and enter 100 minus the 
required duty cycle value. 
Delay
Measured between trigger output and main output. Can be entered as 
absolute delay, phase
°
 or % of period.
Width Agilent 81101A
Range: 10 ns to 999.5 s
(max value: period –10 ns)
Accuracy:
±
5% 
±
250 ps 
Duty cycle: 0.1% to 95% (depends on period and width; 
overprogrammable to 99%)
Delay Agilent 81101A
Fixed delay from 
TRIGGER OUT:
17.0 ns typical
Additional variable range: 0.00 ns to 999.5 s
(max value: period –20 ns)
Accuracy:
±
5% 
 ±
1 ns










