User`s guide
User’s and Service Guide 6-21
Service
Troubleshooting and Block Diagram
Troubleshooting and Block Diagram
This section contains information on troubleshooting the test set to the 
assembly level only. By following these procedures you should be able to 
determine whether the power supply, front panel, or main switch board 
needs replacing. A block diagram is included at the end of this section 
as an aid in troubleshooting.
Theory of operation information can be found in the next section of this 
manual.
General Troubleshooting Notes
WARNING Always turn the instrument power off before removing or 
installing an assembly.
CAUTION If you need to disassemble the instrument, be sure to work at an 
antistatic workstation and use a grounded wrist strap to prevent 
damage from electrostatic discharge (ESD). See Figure 1-3 on page 1-7.
Troubleshooting Power Supply Problems
Turn the instrument on. Check the condition of the LCD on the front 
panel:
Step 1. If the LCD is off, check the main fuse located in the power supply filter 
at the rear of the instrument and replace if necessary.
Step 2. If the LCD is still off, there is still a possibility that the power supply is 
not supplying the necessary +24V, +12V, and +5V to the main board.
Step 3. If the LCD is still off, check the cable between the main board and front 
panel board.
Step 4. Finally, disconnect the DC power cable from the power supply to the 
main switch board and measure the voltages. They should be +24V, 
+12V, and +5V. If not, replace the power supply.










