Specifications
3-2
Testing Performance
This chapter explains how to verify correct oscilloscope operation and 
perform tests to ensure that the oscilloscope meets the performance 
specifications.
To completely test and troubleshoot the mixed-signal oscilloscope, you 
will create and use a test connector accessory, as described in this 
chapter.
• The test connector makes it easy for you to connect the oscilloscope 
probes to function generators and measurement equipment with 
minimum electrical distortion. 
• The connector is used in the digital channel threshold accuracy test.
Let the Equipment Warm Up Before Testing For accurate test results, let 
the test equipment and the oscilloscope warm up 30 minutes before testing.
Verifying Test Results During the tests, record the readings in the 
Performance Test Record at the end of this chapter for your oscilloscope. To 
verify whether a test passes, verify that the reading is within the limits in the 
Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure given in Chapter 4. 
If the User Cal procedure does not correct the problem, refer to Chapter 4, Calibrating 
and Adjusting. 