User`s guide
3-46
Triggering the Oscilloscope
To use TV triggering
Triggering on Line 71 
Line Numbers per Field for Each TV Standard 
TV Standard Field 1 Field 2 Alt Fld 
NTSC 1 to 263 1 to 262 1 to 262 
PAL 1 to 313 314 to 625 1 to 313 
PAL-M 1 to 263 264 to 525 1 to 263
SECAM 1 to 313 314 to 625 1 to 313 
GENERIC 1 to 1024 1 to 1024 1 to 1024 (Vertical) 
Line Number Represents Count
In Generic mode, the line number represents the number of a count instead of a real 
line number. This is reflected in the label in the softkey changing from Line to Count. 
In the Mode softkey selections, Line:Field 1, Line:Field 2 and Count:Vertical are used 
to indicate where the counting starts. For an interlaced TV system, the counting 
starts from the rising edge of the first vertical serration pulse of Field 1 and/or Field 2. 
For a non-interlaced TV system, the counting starts after the rising edge of the 
vertical sync pulse.