Triggering the Oscilloscope
To use TV triggering
If a more detailed analysis is required, then only one color field should be 
selected to be the trigger. You can do this by using the trigger Holdoff softkey 
in the trigger Mode/Coupling menu.  Using the holdoff settings shown in the 
following table, the oscilloscope will now trigger on color field 1 OR color field 
3 when Field 1 is selected.  This is known as odd field selection.  Even fields will 
be selected with Field 2.
Holdoff Settings 
Triggering on Color Field 1 using Holdoff (single-shot) 
Video Standard Fields/Picture Holdoff Range 
NTSC 4 33.5 ms to 50.0 ms 
PAL 8 80.7 ms to 120 ms 
SECAM 4 40.4 ms to 60 ms 
PAL-M 8 80.4 ms to 120 ms 