User`s guide
5-28
Making Measurements
FFT Measurement
2 Press the More FFT softkey to display additional FFT settings.
Scale allows you to set your own vertical scale factors for FFT expressed in 
dB/div (decibels/division). Press the Scale softkey, then turn the Entry knob 
to rescale your math function.
Offset allows you to set your own offset for the FFT. The offset value is in dB 
and is re pr es en te d b y t he ce nte r gra tic ul e o f t he di sp lay. Pre ss th e Offset s of tk ey, 
then turn the Entry knob to change the offset of your math function.
Window selects a window to apply to your FFT input signal:
Hanning - window for making accurate frequency measurements or for 
resolving two frequencies that are close together.
Flat Top - window for making accurate amplitude measurements of frequency 
peaks.
Rectangular - good frequency resolution and amplitude accuracy, but use only 
where there will be no leakage effects. Use on self-windowing waveforms 
such as pseudo-random noise, impulses, sine bursts, and decaying sinusoids.
3 To make cursors measurements, press the Cursors key and set the Source 
softkey to 
Math.
Use the X1 and X2 cursors to measure frequency values and difference between 
two frequency values (∆X). Use the Y1 and Y2 cursors to measure amplitude 
in dB and difference in amplitude (∆Y).
FFT Scale FFT Offset
Return to 
previous menu
Window
Scale and offset considerations
If you do not manually change the FFT scale or offset settings, when you turn the 
horizontal sweep speed knob, the span and center frequency settings will 
automatically change to allow optimum viewing of the full spectrum. If you do 
manually set scale or offset, turning the sweep speed knob will not change the span 
or center frequency settings, allowing you see better detail around a specific 
frequency. Pressing the FFT Preset softkey will automatically rescale the waveform 
and span and center will again automatically track the horizontal sweep speed 
setting.