User`s guide
3-10
Triggering the Oscilloscope
External Trigger Input
2Press the External softkey to display the external trigger probe menu. 
Probe Attenuation Turn the Entry knob to set the attenuation factor 
displayed in the Probe softkey for the connected probe. The attenuation factor 
can be set from 0.1:1 to 1000:1 in a 1-2-5 sequence.
For 2-channel oscilloscopes that have the Ext Trigger connector on the front 
panel, when you connect an AutoProbe self-sensing probe, the oscilloscope will 
automatically configure your probe to the correct attenuation factor.
The probe correction factor must be set properly for measurements to be made 
correctly.
Probe Units Press the Units softkey to select the proper measurement unit 
for the connected probe. Select Volts for a voltage probe and select Amps for a 
current probe. Measurement results, channel sensitivity, and trigger level will 
reflect the measurement units you have selected.
Input Impedance The 54620-series oscilloscope external trigger input 
impedance is always 1 MΩ. The 54640-series oscilloscope external trigger input 
impedance can be set to either 1M Ohm or 50 Ohm by pressing the Imped softkey. 
• 50 Ohm mode matches 50-ohm cables commonly used in making high 
frequency measurements. This impedance matching gives you the most 
accurate measurements since reflections are minimized along the signal 
path.
• 1M Ohm mode is for use with probes and for general-purpose measurements. 
The higher impedance minimizes the loading effect of the oscilloscope on 
the circuit under test.
For 2-channel oscilloscopes that have the Ext Trigger connector on the front 
panel, when you connect an AutoProbe self-sensing probe, the oscilloscope will 
automatically configure your probe to the correct impedance.
Attenuation 
factor
Measurement 
units
Input impedance 
(54640-series)
Return to 
previous menu