User`s guide

6-24
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
NOTE Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
NOTE The thru in most calibration kits is defined with zero length. The correction
will not work properly if a non-zero length thru is used, unless the calibration
kit is modified to change the defined thru to the length used. This is
important for measurements of non-insertable devices (devices having ports
that are both male or both female). The modified calibration kit must be
saved as the user calibration kit, and the softkey must be
selected before the calibration is started.
16.To measure the standard, when the trace has settled, press:
or
or
The analyzer underlines the softkey label after it makes each measurement.
17.Press and select from the following two options:
If you will be measuring devices with a dynamic range less than 90 dB, press:
If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Connect impedance-matched loads to the test ports. Include the adapters that you
would include for your device measurement.
NOTE If you will be measuring highly reflective devices such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard.
b. Activate at least four times more averages than desired during the device
measurement.
c. Press
.
d. Return the averaging to the original state of the measurement, and press
.
USER KIT
FWD TRANS THRU
REV TRANS THRU
FWD MATCH THRU
REV MATCH THRU
STANDARDS DONE
ISOLATION
OMIT ISOLATION
Cal
RESUME CAL SEQUENCE
ISOLATION FWD or REV
ISOL'N STD DONE
Cal
RESUME CAL SEQUENCE