Operating Concepts
Measurement Calibration
In this case, omitting isolation is better than measuring the isolation standards without
increasing the averaging factor.
Figure 7-37 Isolation E
XF
Thus there are two sets of error terms, forward and reverse, with each set consisting of six
error terms, as follows:
Directivity, E
DF
(forward) and E
DR
(reverse)
Isolation, E
XF
and E
XR
Source Match, E
SF
and E
SR
Load Match, E
LF
and E
LR
Transmission Tracking, E
TF
and E
TR
Reflection Tracking, E
RF
and E
RR
The analyzer's test set can measure both the forward and reverse characteristics of the
test device without you having to manually remove and physically reverse the device. A
full two-port error model illustrated in Figure 7-38. This illustration depicts how the
analyzer effectively removes both the forward and reverse error terms for transmission
and reflection measurements.