User`s guide

7-55
Operating Concepts
Calibration Routines
Full Two-Port Calibration (ES Models Only)
The full two-port calibration, activated by pressing the softkey within the
calibrate menu, provides directivity, source match, load match, isolation, and frequency
response vector error-correction, in both forward and reverse directions, for transmission
and reflection measurements of two-port devices. This calibration provides the best
magnitude and phase measurement accuracy for both transmission and reflection
measurements of two-port devices, and requires an S-parameter test set.
In this type of calibration, both forward and reverse measurements must be made. You
have the option of setting the ratio of the number of forward (or reverse) sweeps versus the
number of reverse (or forward) sweeps. To access this function, press
and enter the number of sweeps desired.
TRL*/LRM* Two-Port Calibration
The TRL*/LRM* two-port calibration, activated by pressing the
softkey within the calibration menu, provides the ability to make calibrations using the
TRL or LRM method. True TRL/LRM calibration is available on analyzers equipped with
Option 400, the Four Sampler Test Set. For more information, refer to “TRL*/LRM*
Calibration (ES Models Only)” on page 7-66.
FULL 2-PORT
Cal
MORE
TESTSET SW
TRL*/LRM* 2-PORT