User`s guide

7-85
Operating Concepts
Differences between 8753 Network Analyzers
Differences between 8753 Network Analyzers
Table 7-3 Comparing the 8753A/B/C/D
Feature 8753A 8753B 8753C 8753D 8753D
Opt. 011
Fully integrated measurement system
(built-in test set)
No No No Yes No
Test port power range (dBm)
a
a. For this network analyzer, the feature is dependent on the test set being used.
a
a
+10 to
85
a
Auto/manual power range selecting No No No Yes No
Port power coupling/uncoupling No No No Yes No
Internal disk drive No No No Yes Yes
Precision frequency reference (Option 1D5) No No No Yes Yes
Frequency range - low end (in kHz) 300 300 300 30
30/300
b
b. 300 kHz to 3 GHz without Option 006; 30 kHz to 6 GHz, with Option 006.
Ext. frequency range to 6 GHz (Option 006) No Yes Yes Yes Yes
75 system impedance (Option 075)
a
a
a
Yes
a
TRL*/LRM* correction No No No Yes Yes
Power meter calibration No Yes Yes Yes Yes
Interpolated error correction No Yes Yes Yes Yes
Maximum error corrected measurement points 801 1601 1601 1601 1601
Segmented error correction in frequency list mode No No Yes Yes Yes
Color CRT No No Yes Yes Yes
Test sequencing No Yes Yes Yes Yes
Automatic sweep time No Yes Yes Yes Yes
External source capability No Yes Yes Yes Yes
Tuned receiver mode No Yes Yes Yes Yes
Printer/plotter buffer No Yes Yes Yes Yes
Harmonic measurements (Option 002) No Yes Yes Yes Yes
Frequency offset mode (mixer measurements) No Yes Yes Yes Yes
dc bias to test device
a
a
a
Yes
a
Interfaces: RS-232, parallel, and DIN keyboard No No No Yes Yes
User-defined preset No No No Yes Yes
Non-volatile memory (in Kbytes) 16 16 16 512 512
Dynamic range: 30 kHz 3 GHz 100 dB 100 dB 100 dB
110 dB
c
c. 90 dB from 30 kHz to 50 kHz; 100 dB from 300 kHz to 16 MHz.
100 dB
Dynamic range: 3 GHz 6 GHz N/A 80 dB 80 dB 105 dB 110 dB
Real time clock No No No Yes Yes