4 Measurement Tutorial
On the following page, the first figure shows the Fourier Integral of two very 
different pulses: one of broad width (200 
μs); the other narrow (6.7 μs). The 
second figure is an approximation of the shape of the input frequency 
response of the 34410A/11A/L4411A. 
Comparing these two figures along the frequency axis, notice that 
the   spectrum of the narrow pulse significantly exceeds the 
effective bandwidth of the instrument. The net result is a less accurate 
measurement of the narrow, high–frequency pulse.
In contrast, the frequency spectrum of the broad pulse has fallen off 
significantly below the multimeter’s 300 kHz (approximately) band–width, so 
measurements of this pulse are more accurate. 
Reducing the prf increases the density of lines in the Fourier spectrum, and 
increases the portion of the input signal’s spectral energy within the 
multimeter’s bandwidth, which improves accuracy.
In summary, error in rms measurements arise when there is significant input 
signal energy at frequencies above the multimeter’s bandwidth.