2Features and Functions
Configuring Frequency and Period Measurements
For frequency and period measurements, your menu selections are: CONFIGURE, 
GATE TIME, RANGE, AC FILTER, NULL and NULL VALUE.
• CONFIGURE: Allows you to select either FREQUENCY or PERIOD as the primary 
measurement.
• GATE TIME: Allows you to select one of four choices (0.001, 0.01, 0.1, or 1) for 
gate time, in seconds.
• RANGE: Allows you to let the multimeter automatically select the range using 
the autoranging feature (select AUTO), or you may select a fixed range 
(select MANUAL).
• AC FILTER: Allows you to select from three choices (3 HZ : SLOW, 
20 HZ : MEDIUM, or 200 HZ : FAST). The ac filter allows you to trade off 
low–frequency bandwidth versus ac settling time.
• NULL: Allows you to enable (ON) or disable (OFF) the null measurement 
feature, which measures the difference between a stored null value and the 
input signal.
• NULL VALUE: Allows you to view and edit the null value (if enabled). 
Configuring Temperature Measurements
For temperature measurements, your menu selections are: PROBE TYPE, OFFSET 
COMP, AUTO ZERO, INTEGRATION, NULL, NULL VALUE, and UNITS.
• PROBE TYPE: Allows you to select from four choices (RTD–4W, RTD–2W, 
THERMISTOR–2W, or THERMISTOR–4W) of temperature probe type.
• OFFSET COMP (for RTD probes only): Allows you to enable (ON) or disable 
(OFF) the offset compensation feature. With offset compensation enabled, 
the multimeter makes a normal temperature measurement first, followed by 
a second measurement to determine any offset voltage in the input 
circuitry. The resultant displayed measurement corrects for this offset. 
Enabling offset compensation increases measurement time.
• AUTO ZERO: Allows you to enable (ON) or disable (OFF) the auto zero feature 
for 2–wire temperature measurements. This feature subtracts a subsequent 
zero reading from each measurement.
Note that 4–wire temperature measurements are automatically made 
with auto zero always on.
• INTEGRATION: Allows you to set the integration time for the measurement in 
two ways; in power–line cycles (NPLC) or in seconds (APERTURE).