2–wire Resistance Measurements:
4–wire Resistance Measurements
Capacitance
AC Voltage Measurements (all ranges):
AC Current Measurements (all ranges):
Frequency and Period
Range Trigger Delay
(for ≤ 0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
 (for ≥ 1 PLC)
100 Ω 80 μs100 μs130 μs
 1 kΩ 110 μs130 μs160 μs
10 kΩ 130 μs160 μs190 μs
100 kΩ 540 μs670 μs800 μs
   1 MΩ  5.0 ms 6.0 ms 7.5 ms
10 MΩ 
– 1GΩ  60 ms  70 ms  84 ms
Range Trigger Delay
(for ≤ 0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
 (for ≥ 1 PLC)
100Ω 
– 100 kΩ 1.0 ms   1.0 ms 1.5 ms
 1 MΩ 10 ms 10 ms 15 ms
10 MΩ 
– 1GΩ 100 ms  100 ms 100 ms
Trigger Delay
   0 s
AC Filter Trigger Delay 
Slow (3 Hz) 2.5 s
Medium (20 Hz)   625 ms
Fast (200 Hz) 25 ms
AC Filter Trigger Delay 
Slow (3 Hz) 1.66 s
Medium (20 Hz) 250 ms
Fast (200 Hz)  25 ms
Trigger Delay
   1 s