34410A/11A/L4411A User’s Guide 93
Remote Interface Configuration 3
IP Address 
An Internet Protocol (IP) Address is required for all IP and TCP/IP 
communications with the instrument. If DHCP is enabled (factory setting), the 
specified static IP address is not used. However, if the DHCP server fails to 
assign a valid IP address, or if DHCP and Auto–IP are both off, the currently 
configured static IP address will be used.
• Changing the IP address will cause an automatic LAN reset.
• The factory–set default IP Address is “169.254.4.10”.
• If you are planning to use a static IP address on a corporate LAN, contact 
your network administrator to obtain a fixed IP address to be used 
exclusively for your instrument.
• The IP address is stored in non–volatile memory, and does not change when 
power has been off, after a Factory Reset (*RST) command, or after an 
Instrument Preset (SYSTem:PRESet) command.
Subnet Mask
The multimeter uses the Subnet Mask to determine if a client IP address is on 
the same local subnet. When a client IP address is on a different subnet, all 
packets must be sent to the Default Gateway. Contact your network 
administrator to determine if subnetting is being used and for the correct 
Subnet Mask.
• Changing the Subnet Mask setting will cause an automatic LAN reset.
• The default Subnet Mask for the 34410A/11A multimeter is “255.255.0.0”.
• A value of “0.0.0.0” or “255.255.255.255” indicates that subnetting is not 
being used.
• The Subnet Mask is stored in non–volatile memory, and does not change 
when power has been off, after a Factory Reset (*RST) command, or after 
an Instrument Preset (SYSTem:PRESet) command.