Specifications
Chapter 4 4-13
Appendix
DANL Manual Performance Verification
Table 4-10  Agilent E4443A Performance Verification Test Record
Agilent Technologies
Model E4443A Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum
Results 
Measured
Maximum
Measurement 
Uncertainty
Displayed Average 
Noise Level
Note: Enter results with preamp on in the appropriate section based upon 
the ambient temperature when the test was performed. 
Preamp On
 100 kHz (1)________ –159 dBm ±0.07 dB
 199 kHz (2)________ –159 dBm ±0.07 dB
 201 kHz (3)________ –159 dBm ±0.07 dB
 499 kHz (4)________ –159 dBm ±0.07 dB
 501 kHz (5)________ –163 dBm ±0.07 dB
 9.9 MHz (6)________ –166 dBm ±0.07 dB
 10.1 MHz (7)________ –169 dBm ±0.07 dB
 1.0 GHz (8)________ –168 dBm ±0.07 dB
 1.2 GHz (9)________ –167 dBm ±0.07 dB
 2.4 GHz (10)________ –165 dBm ±0.07 dB
 2.6 GHz (11)________ –165 dBm ±0.07 dB
 3.0 GHz (12)________ –165 dBm ±0.07 dB
Preamp Off
 10 kHz (13)________ –134 dBm ±0.07 dB
 99 kHz (14)________ –134 dBm ±0.07 dB
 101 kHz (15)________ –144 dBm ±0.07 dB
 990 kHz (16)________ –144 dBm ±0.07 dB
 1.01 MHz (17)________ –149 dBm ±0.07 dB
 500 MHz (18)________ –154 dBm ±0.07 dB
 1.1 GHz (19)________ –154 dBm ±0.07 dB
 1.3 GHz (20)________ –153 dBm ±0.07 dB
 2.0 GHz (21)________ –153 dBm ±0.07 dB
 2.4 GHz (22)________ –153 dBm ±0.07 dB