Specifications
12
Noise figure measurements 
using a mixer as the DUT
When a down conversion is included 
in the noise figure measurement, for
example measuring the noise figure
of a mixer, there are some additional
setups to consider. For this example
let us use a mixer as a down-converter
with an IF at 70MHz, LO at 1GHZ
and both RF sidebands are used, 
930 and 1070 MHz (DSB): 
• The measurement, as well as 
calibration, is made at the 
IF frequency.
• When an IF frequency is chosen, 
it is a good idea to keep the 
frequency as low as possible in 
order to avoid large differences in 
ENR values between the upper 
and lower sidebands when using 
DSB mode. This is because it is 
the ENR value at the LO that is 
used in the measurement 
(compromise since it is centered 
between the two sidebands) 
• Since this device has some loss, it 
is recommended that the internal 
preamp be used. 
• Compensate for two sidebands by 
selecting double side band. Any 
broadband noise in the LO will 
directly affect results. This can 
be solved by either a high pass or 
low pass filter at the LO port that 
removes the noise at the IF 
frequency. Place an IF filter at the 
input of the spectrum analyzer to 
remove LO feed through. Usually 
mixers have around 20 dB of 
isolation between the LO-IF port 
so the high powered LO will 
seriously affect results.
Figure 13.
Setup for 
measuring a mixer 
Instructions Keystrokes
Set up the PSA for down conversion  {Meas Setup} {Int Preamp On
}
measurements. It is recommended that the  [Mode Setup] {DUT Setup…} {Down Conv}
internal preamp be used when measuring 
devices that have low gain. 
Set up the source for +7 dBm at 1 GHz.  On E4438C press [Frequency] [1] {GHz}
[Amplitude] [7] {dBm} [RF On]. 
Set up the LO frequency (Figure 13).  Scroll to “Ext LO frequency” using tab keys 
then enter [1] {GHz}. Tab to “sideband” and 
choose “DBS”. 
Set up the fixed IF frequency.  [Frequency] {Freq Mode} {Fixed}
{Fixed Frequency} [70] {MHz} 
Calibration: connect the noise source to  [Meas Setup] {Calibrate} {Calibrate}
the input of the PSA. 
Measure the DUT: Connect the mixer IF (I)  To add more averaging, press [Meas Setup] 
port to the PSA, the LO (L) port to the signal  then {Avg Number On
). 
source and the RF (R) to the noise source. 
Perform measurements on mixers.