Specifications

C
Copyright 2012 Agilent Technologies
Change 8
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test (20
cm AIR LINE (0 m)).
20 cm AIR LINE (0 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Frequency
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
1.1 MHz
1 V
C.V. - 0.0316 pF
C.V. + 0.0316 pF
0.0072 pF
-0.0023
0.0023
0.0005
1.1 MHz
2 V
C.V. - 0.0282 pF
C.V. + 0.0282 pF
0.0072 pF
-0.0021
0.0021
0.0005
2 MHz
1 V
C.V. - 0.0303 pF
C.V. + 0.0303 pF
0.0076 pF
-0.0022
0.0022
0.0005
2 MHz
2 V
C.V. - 0.04505 pF
C.V. + 0.04505 pF
0.0076 pF
-0.0033
0.0033
0.0005
5 MHz
1 V
C.V. - 0.0518 pF
C.V. + 0.0518 pF
0.0110 pF
-0.0038
0.0038
0.0008
5 MHz
2 V
C.V. - 0.04685 pF
C.V. + 0.04685 pF
0.0110 pF
-0.0035
0.0035
0.0008
10 MHz
20 mV
C.V. - 0.2182 pF
C.V. + 0.2182 pF
0.0230 pF
-0.0161
0.0161
0.0016
10 MHz
100 mV
C.V. - 0.1121 pF
C.V. + 0.1121 pF
0.0230 pF
-0.0083
0.0083
0.0016
10 MHz
1 V
C.V. - 0.1121 pF
C.V. + 0.1121 pF
0.0230 pF
-0.083
0.0083
0.0016
10 MHz
2 V
C.V. - 0.0988 pF
C.V. + 0.0988 pF
0.0230 pF
-0.0073
0.0073
0.0016
20 MHz
1 V
C.V. - 0.2392 pF
C.V. + 0.2392 pF
0.0690 pF
-0.0177
0.0177
0.0049
20 MHz
2 V
C.V. - 0.2739 pF
C.V. + 0.2739 pF
0.0690 pF
-0.0203
0.0203
0.0049
30 MHz
1 V
C.V. - 0.5201 pF
C.V. + 0.5201 pF
0.1500 pF
-0.0385
0.0385
0.0105
30 MHz
2 V
C.V. - 0.4938 pF
C.V. + 0.4938 pF
0.1500 pF
-0.0365
0.0365
0.0105