Specifications

C
Copyright 2012 Agilent Technologies
Change 11
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test (20
cm AIR LINE (1 m)).
20 cm AIRLINE (1 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Frequency
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
10 MHz
1 V
C.V. - 0.2022 pF
C.V. + 0.2022 pF
0.0230 pF
-0.0150
0.0150
0.0016
10 MHz
2 V
C.V. - 0.1889 pF
C.V. + 0.1889 pF
0.0230 pF
-0.0140
0.0140
0.0016
30 MHz
1 V
C.V. - 0.7903 pF
C.V. + 0.7903 pF
0.1500 pF
-0.0585
0.0585
0.0105
30 MHz
2 V
C.V. - 0.7640 pF
C.V. + 0.7640 pF
0.1500 pF
-0.0565
0.0565
0.0105
Change 12
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test
(100pF STANDARD (2 m)).
100 pF STANDARD (2 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
1 V
C.V. - 0.220 pF
C.V. + 0.220 pF
0.0250 pF
-0.0022
0.0022
0.0002
2 V
C.V. - 0.285 pF
C.V. + 0.285 pF
0.0250 pF
-0.0029
0.0029
0.0002