Specifications
4-3
Calibrating and Adjusting
WARNING  HAZARDOUS VOLTAGES !
Read the safety summary at the back of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with 
the protective covers removed.  Only trained service personnel who are aware 
of the hazards involved should perform the maintenance.  Whenever possible, 
perform the procedures with the power cord removed from the oscilloscope. 
CAUTION  REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies with power applied 
to the oscilloscope. Otherwise, damage to the oscilloscope can occur. 
CAUTION  USE EXTERNAL FAN TO REDUCE TEMPERATURE !
When you must operate the oscilloscope with its cover removed, use an 
external fan to provide continuous air flow over the samplers. Air flow over 
the samplers is reduced when the cover is removed, which leads to higher than 
normal operating temperatures. Have the fan blow air across the system PC 
board where the heat sinks are located. 
CAUTION  AVOID DAMAGE TO ELECTRONIC COMPONENTS !
Electrostatic discharge (ESD) can damage electronic components.  When you 
use any of the procedures in this chapter, use proper ESD precautions. As a 
minimum, place the oscilloscope on a properly grounded ESD mat and wear a 
properly grounded ESD strap.
service.book Page 3 Wednesday, August 28, 2002 3:25 PM