User`s guide
3-6
Programming over GPIB
Lockout
Lockout
With GPIB, the instrument is placed in the lockout mode by sending the local 
lockout command (LLO). The instrument can be returned to local by sending 
the go-to-local (GTL) command to the instrument.
Bus Commands 
The following commands are IEEE 488.1 bus commands (ATN true). IEEE 
488.2 defines many of the actions which are taken when these commands are 
received by the instrument. 
Device Clear 
The device clear (DCL) or selected device clear (SDC) commands clear the 
input and output buffers, reset the parser, and clear any pending commands. If 
you send either of these commands during a digitize operation, the digitize 
operation is aborted. 
Interface Clear (IFC) 
The interface clear (IFC) command halts all bus activity. This includes 
unaddressing all listeners and the talker, disabling serial poll on all devices, and 
returning control to the system controller. 