Safety Summary When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Agilent Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for the operator. n Instrument operates abnormally. n Instrument emits abnormal noise, smell, smoke or a spark-like light during the operation.
Caution Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument's Function Reference. Electrostatic discharge(ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures or cables are being connected or disconnected. Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground.
Agilent 43521A Downconverter Unit Operation Manual Agilent Part No.
Notice The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies Japan, Ltd. Agilent Technologies Japan, Ltd.
Manual Printing History The manual's printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The manual part number changes when extensive technical changes are incorporated.
Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with speci c WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer's failure to comply with these requirements.
Dangerous Procedure Warnings Warnings , such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. Warning Dangerous voltages, capable of causing death, are present in this instrument. Use extreme caution when handling, testing, and adjusting this instrument.
Typeface Conventions Bold Italics Computer 4HARDKEYS5 NNNNNNNNNNNNNNNNNNNNNNNNNN SOFTKEYS vi Boldface type is used when a term is de ned. For example: icons are symbols. Italic type is used for emphasis and for titles of manuals and other publications. Italic type is also used for keyboard entries when a name or a variable must be typed in place of the words in italics. For example: copy lename means to type the word copy, to type a space, and then to type the name of a le such as file1.
Certi cation Agilent Technologies certi es that this product met its published speci cations at the time of shipment from the factory. Agilent Technologies further certi es that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution's calibration facility, or to the calibration facilities of other International Standards Organization members.
Safety Symbols General de nitions of safety symbols used on equipment or in manuals are listed below. Instruction manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instruction manual. Alternating current. Direct current. On (Supply). O (Supply). In position of push-button switch. Out position of push-button switch. Frame (or chassis) terminal. A connection to the frame (chassis) of the equipment which normally include all exposed metal structures.
Contents 1. Unpacking and Checking the Contents of the Shipment Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Contents of the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . 2. Connecting 43521A Notes on Installation . . . . . . . . . . . . . . . . . . Requirements for Operating Environment . . . . . . . Providing clearance to dissipate heat at installation site Instruction for Cleaning . . . . . . . . . . . . . . .
5. Speci cations Speci cations of the 43521A . . . . . . . . . . . . . . . . . . . . . . . . RF Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . Non-Operating Environment . . . . . . . . . . . . . . . . . . . . . . Others . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System Speci cations . . . . . . . . . . . . . . . . . . . . . . . . . . . 1. When Frequency Band is 10 MHz to 3 GHz . . . . . . . . . .
Figures 2-1. 2-2. 2-3. 3-1. 3-2. 4-1. A-1. Connections on the Front Panel . Connections on the Rear Panel . Optional Power Cables . . . . . Front Panel . . . . . . . . . . rear panel . . . . . . . . . . . Rack Mount Kit Installation . . Serial Number Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unpacking and Checking the Contents of the Shipment 1 This chapter describes how to check the shipping container of your 43521A ( Downconverter Unit), the appearance of the product, and the contents of the shipment. Incoming Inspection When you receive the product, inspect the shipping container for damage.
2 Connecting 43521A The 43521A (Downconverter Unit) is used as part of the 4352S (VCO/PLL Signal Test System). Using the 43521A expands the maximum frequency of the 4352S to 12.6 GHz. This chapter describes how to make connection for the 4352S when associated with the 43521A. How to use the entire 4352S including the 43521A is described in Chapter 5 of the 4352B Function Reference, while a quick guide to diagnosis is provided in Appendix F of that document.
Conditions Rear 180 mm Side 60 mm Instruction for Cleaning To prevent electrical shock, disconnect the 43521A's power cable from the power outlet before cleaning. To clean the exterior of the 43521A, wipe the surfaces with a dry cloth or a soft cloth that is soaked with water and wrung tightly, without undue pressure. Do not attempt to clean the 43521A internally.
Connections on the Front Panel Make connections as shown in Figure 2-1. For how to connect a DUT (Device Under Test), refer to Chapter 5 of the 4352B Function Reference. Figure 2-1. Connections on the Front Panel No. Name 1 N-N cable (18 cm x 2) 2 N-N cable Table 2-1.
Connections on the Rear Panel Make connections as shown in Figure 2-2. Figure 2-2. Connections on the Rear Panel No. Name Table 2-2.
Power Requirements Ratings of the Power Source The 43521A requires the following power source: Voltage : 90 to 132 Vac or 198 to 264 Vac Frequency : 47 to 63 Hz Power : 100 VA maximum Power Cable In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame. The type of power cable shipped with each instrument depends on the country of destination.
Figure 2-3.
3 Front Panel and Rear Panel This chapter explains the front and rear panels of the 43521A (Downconverter Unit). How to use the entire 4352S including the 43521A is described in Chapter 5 of the 4352B Function Reference. Related sample programs are found in \Frequency Transient Measurement" in Chapter 12 of the 4352B GPIB Programming Manual. Front Panel Figure 3-1 shows the con guration of the front panel. Figure 3-1.
1. RF Input Connector Receives the RF output signal from the device under test (signal to be measured). INSTALLATOIN CATEGORY I 2. RF Output Connector The RF signal (signal to be measured) inputted to the RF input connector is outputted from this connector, and sent to the RF input connector of the 4352B (VCO/PLL Signal Analyzer) through the N-N cable.
Rear Panel Figure 3-2 shows the con guration of the rear panel. Figure 3-2. rear panel 1. 40 MHz Input Terminal To improve the frequency accuracy of the 600-MHz signal inside the 43521A, the 40-MHz signal from the rear panel of the 4352B (VCO/PLL Signal Analyzer) is inputted to this connector to provide phase locking for synchronizing the 600-MHz signal to the 40-MHz signal. 2. I/O Port Receives the digital signal from the I/O port connector on the rear panel of the 4352B.
Replacing the Fuse Select a proper fuse according to the Table 3-1. Table 3-1. Rating of Fuse Rating Part Number 3A 250Vac UL/CSA type Time delay 2110-0381 For ordering the fuse, contact your nearest Agilent Technologies Sales and Service O ce. Replacing the Fuse Lever a small minus screwdriver to dismount the fuse holder next to the AC line receptacle on the rear panel. Caution Replace the fuse in the fuse holder. When replacing the fuse, be sure to check that the rating and the type are correct.
4 Rack/Handle Installation Rack/Handle Installation This instrument can be rack-mounted and used as a component in a measurement system. Figure 4-1 shows how to rack-mount the instrument. Option Table 4-1. Rack Mount Kits Description 1CN 1CM 1CP Front Handle Kit Rack Mount Kit Rack Mount & Handle Kit Agilent Part Number 5063-9226 5063-9212 5063-9219 Figure 4-1.
Option 1CN Handle Kit Option 1CN includes a pair of handles and accessories to attach them to the instrument. Installing the Handle 1. Remove the cosmetic panels 1 a xed to both sides of the frame on the front panel. (See Figure 4-1.) 2. Attach the handles 3 to both sides of the instrument using the screws provided. 3. Attach the trim strips 4 to the handles. Option 1CM Rack Mount Kit Option 1CM includes a pair of anges and accessories to mount the instrument to an EIA-compliant standard rack (482.
5 Speci cations The speci cations mean the performance standards and operation limits. They are the 43521A's performance guaranteed, unless otherwise speci ed, over the temperature range of 0 C to 40 C and after a warm-up time of at least 30 minutes has elapsed from power-on. Information provided as typical, reference, or approximate, though not speci cations, is intended to serve as reference data for e cient operation of the 43521A.
Operating Environment Temperature Humidity Altitude Warm up time 0 C to 40 C 15% to 95% RH 0 to 2000 m 30 minutes ::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: ::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: :::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: ::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: Non-Operating Environment Temperature Humidity :::::::::::::::::::::::::::::::::::::::::::::::::
System Speci cations The following provide the speci cations of the system in which the 43521A is connected to the 4352B (VCO/PLL Signal Analyzer) and the external signal source. It is assumed that the 43521A is locked by the 40 MHz signal from the 4352B. All values, except for the RF power measurement in 2-1. Tester Mode, are reference values. Note that same speci cations as those of the 4352B are not shown. 1.
Table 5-2.
A Manual Changes Introduction This appendix contains the information required to adapt this manual to earlier versions or con gurations of the 43521A than the current printing date of this manual. The information in this manual applies directly to the 43521A serial number pre x listed on the title page of this manual. Manual Changes To adapt this manual to your 43521A see Table A-1 and make all the manual changes listed opposite your instrument's serial number and rmware version.
Serial Number Agilent Technologies uses a two-part, ten-character serial number that is stamped on the serial number plate (Refer to Figure A-1) attached to the rear panel. The rst ve characters are the serial pre x and the last ve digits are the su x. Figure A-1.
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. 11/29/99 United States: Agilent Technologies Test and Measurement Call Center P.O.