User`s guide
General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 81
Calibration Procedure Using 8480 Series Power Sensors and 
N8480 Series Power Sensors (with Option CFT)
The following procedure describes how you calibrate the power meter with 
the 8480 Series power sensors and N8480 Series power sensors with 
Option CFT.
There are a variety of methods to connect the power sensors to the power 
meter depending on the model of power sensor you are using. Refer to 
Table 2- 5 on page 82 for details on connecting different power sensor 
models.
1 Press ,  and . 
Verify the reference calibration factor of your power sensor with that 
displayed under  . The value shown is obtained from the sensor 
calibration table (if one is selected), otherwise it is the last value set or 
the default of 100%. If the value is incorrect, press  . The 
reference calibration factor pop-  up window will be displayed. Use the 
numeric keypad to enter the required value in the pop- up window.
2 Confirm your choice by pressing  .
3 Press the   to start the calibration routine. The Calibrating pop- up is 
then displayed (the power meter automatically turns on the POWER 
REF output).
NOTE
V8486A and W8486A sensors
For most 8480 Series sensors, the correct (A type or D type) linearity correction table is 
automatically selected. However, for the V8486A and W8486A sensors, the automatic 
selection must be overridden and the D type correction selected. Subsequent connection of 
another A type sensor will result in a warning message stating that “Linearity Override May 
be Required”.
To select the linearity type to be applied,
• Press  ,   and   to select either   or  .
Tab les
Linearity
Atyp
Dtyp
2 of 3
Ref CFs
Ref CFs
Ref CFs
%
Cal










