User`s guide
6 Using N8480 Series Power Sensors
148 N1913/1914A EPM Series Power Meters User’s Guide
Introduction
The N8480 Series power sensors are replacement for the 8480 Series power 
sensors (excluding the D- model sensors) with build- in 
E
lectrically 
E
rasable 
P
rogrammable 
R
ead-
O
nly 
M
emory
 (
EEPROM)
1
.
The N8480 Series power sensors are used for measuring the average power 
supplied by RF or microwave source or a device- under- test (DUT). The N8480 
Series power sensors place a 50 
Ω
 load on the RF or microwave source. The 
power meter indicates the power dissipated in this load in W or dBm.
The N8480 sensors (excluding Option CFT) measure power levels from –35 
dBm to +20 dBm (316 nW to 100 mW), at frequencies from 10 MHz to 33 
GHz and have two independent power measurement path (upper and lower 
range).
Table 6 - 1 Power range in the Range setting 
Meanwhile, the N8480 sensors with Option CFT only measure power levels 
from –30 dBm to +20 dBm (1 µW to 100 mW) in single range. 
Similar to the E- Series power sensors, the N8480 Series power sensors are 
also equipped with EEPROM to store sensor’s characteristics such as model 
number, serial number, linearity, temperature compensation, calibration 
factor
1
 and so forth. However, the calibration factor table stored in EEPROM 
is not applicable for N8480 Series power sensors with Option CFT, and 
require that you use default calibration tables or manually enter the required 
correction factors. Likewise, they cannot be used to make peak or time gated 
measurements.
Please refer to the documentation supplied with your Agilent N8480 Series 
power sensors for specification and calibration information.
1 
The calibration factor table stored in the EEPROM is not applicable for N8480 Series sensors with 
Option CFT.
2 
Recommended for pulse signals measurement with period of more than one second.
Sensor Range Setting Lower Range Upper Range
N8481/2/5/7/8A 
excluding Option CFT
AUTO (Default) –35 dBm to –1 dBm –1 dBm to +20 dBm
LOWER –35 dBm to –1 dBm -
UPPER
2
- –30 dBm to +20 dBm










