N8480 Series Power Sensors (excluding Option CFT)
Power sensors have small errors in their response over frequency. The 
response of each sensor is measured during manufacture (and during 
periodic calibration). With N8480 Series power sensors (excluding Option 
CFT), the resulting frequency compensation information is written into 
EEPROM. This allows the frequency and calibration data to be downloaded 
to the power meter automatically.
Using calibration factors enables you to achieve improved measurement 
accuracy. This section describes making continuous wave measurements 
using the N8480 Series power sensors (excluding Option CFT).
Making a measurement requires the following steps:
1 Zero and calibrate the power meter/sensor combination.
2 Set the frequency for the signal you want to measure.
3 Make the measurement.
Procedure 
1 First, zero and calibrate the power meter/sensor combination.
2 Ensure the power sensor is disconnected from any signal source.Press 
 and the channel   softkey. The Zeroing pop- up is displayed. 
3 Connect the power sensor to the POWER REF output.Press   and 
the channel   softkey to start the calibration routine. The Calibrating 
pop- up is then displayed.
Zero
Cal
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as 
follows:
1 Connect the power sensor to the POWER REF output.
2 Press   and  . (For dual channel meters, press 
,  or   as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B