User`s guide
6 Using N8480 Series Power Sensors
154 N1913/1914A EPM Series Power Meters User’s Guide
N8480 Series Power Sensors with Option CFT
Power sensors have small errors in their response over frequency. The 
response of each sensor is measured during manufacture (and during 
periodic calibration). The calibration factor table written into EEPROM is 
not applicable for N8480 Series power sensors with Option CFT. Therefore, 
the response of each sensor is measured during manufacture (and during 
periodic calibration) and the resulting frequency compensation information 
is supplied in the form of calibration factors. The EPM Series power 
meters provide two methods of using the calibration factors:
• inputting the individual calibration factor for a frequency prior to 
making the measurement, or
• using sensor calibration tables.
If you are making most of your measurements at a single frequency, or in 
a narrow range of frequencies, entering a specific calibration factor is a 
more effective method. Only a minimal amount of data entry is required.
However, if you are making measurements on a wide range of signal 
frequencies, a sensor table is more effective as you only need to enter the 
frequency of the signal you are measuring. The power meter automatically 
selects and applies the calibration factor from the selected table.
Frequency Specific Calibration Factors
This section shows you how to make a measurement using the calibration 
factor for the frequency of the signal you want to measure.
Tip This method is best suited to making several measurements at one 
frequency as you need only enter a small amount of data.
Using this method requires the following steps:
1 Zero and calibrate the power meter/sensor combination.
2 Set the calibration factor value for the frequency of the signal you 
want to measure.
3 Proceed to make the measurement.










