6 Using N8480 Series Power Sensors
156 N1913/1914A EPM Series Power Meters User’s Guide
Now, set the sensor calibration factor for the frequency of the signal to 
measure.
9 Check the current reference calibration factor setting by pressing 
,  . The value is displayed on the Cal Fac field.
10 To change the settings, use the   and   keys to highlight the Cal Fac 
value field and press   to display the Cal Factor pop- up. Use the 
numeric keypad to enter the required value in the Cal Factor pop- up 
window.
Figure 6-6 Calibration factor pop-up window
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as 
follows:
1 Connect the power sensor to the POWER REF output.
2 Press   and  . (For dual channel meters, press 
,   or   as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
Offset
NOTE
Does this setting match the value for the sensor?
The power sensor reference calibration factor can normally be found above the calibration 
factors table on the power sensor body.