User`s guide
Using N8480 Series Power Sensors 6
N1913/1914A EPM Series Power Meters User’s Guide 159
Sensor Calibration Tables
This section describes how to use sensor calibration tables. Sensor 
calibration tables store the measurement calibration factors, for a power 
sensor model or for a specific power sensor, in the power meter. They are 
used to correct measurement results.
Use sensor calibration tables when you want to make power measurements 
over a range of frequencies using one or more power sensors. 
The N1913/1914A EPM Series power meters are capable of storing 20 
sensor calibration tables, each containing up to 80 frequency points. The 
power meter is supplied with a set of nine predefined sensor calibration 
tables plus a “100%” default table. The data in these tables is based on 
statistical averages for a range of Agilent Technologies power sensors. Your 
own sensor will most likely differ from the typical to some degree. If you 
require best accuracy, create a custom table for each sensor you use as 
shown in“Editing/Generating Sensor Calibration Tables” on page 163.
To use calibration factor tables,
1 Select the sensor calibration table to be applied to a channel. 
2 Zero and calibrate the power meter. The reference calibration factor 
used during the calibration is automatically set by the power meter 
from the sensor calibration table. 
3 Specify the frequency of the signal you want to measure. The 
calibration factor is automatically set by the power meter from the 
sensor calibration table. 
4 Make the measurement.