Using N8480 Series Power Sensors 6
N1913/1914A EPM Series Power Meters User’s Guide 163
Editing/Generating Sensor Calibration Tables
To help achieve the best accuracy in your measurement you can enter the 
values supplied for the sensors you are using by editing the installed 
sensor calibration tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any 
additional tables. However, you can edit or delete the content of each 
table. If you need another table you should edit and rename one of the 
tables. Each calibration table can contain a maximum of 80 frequency 
points.
To view the calibration tables currently stored in the power meter, press 
, ,  . The  Sensor Tbls screen is displayed as 
shown in Figure 6- 8.
Table 6 - 3 Installed power sensor models
1
DEFAULT is a sensor calibration table where the reference calibration factor and calibration 
factors are 100%. This sensor calibration table can be used during the performance testing of the 
power meter.
NOTE
Predefined sensor calibration factor table stored in power meter is not applicable for 
Agilent N8480 Series power sensors with Option CFT. Therefore, users are required to 
create a new sensor calibration table for the sensors when a sensor calibration table is 
needed.
Table Sensor Model Table Sensor Model
0DEFAULT
1
5 8485A
1 8481A 6 R8486A
2 8482A
2
7 Q8486A
3 8483A 8 R8486D
4 8481D 9 8487A
Tab les
Sensor Cal Tables