7 Using U2000 Series USB Power Sensors
176 N1913A/1914A EPM Series Power Meters User’s Guide
Procedure
Set the range as follows:
1 Press 
. On dual channel meters, select the required channel. 
2 The  Range: setting field will be selected. Press   to display the 
Range pop- up.
3 Use the   and   keys to select the required setting.
4 Press   to complete the entry.
Measurement Considerations
While autoranging is a good starting point, it is not ideal for all 
measurements. Signal conditions such as crest factor or duty cycle may 
cause the power meter to select a range which is not the optimum 
configuration for your specific measurement needs. Signals with average 
power levels close to the range switch point require you to consider your 
needs for measurement accuracy and speed. For example, using an 
U2000/1/4A sensor, where the range switch point is –10 ± 1 dBm in a 
pulsed signal configured as follows: 
The calculated average power is –12 dBm.
Characteristic Value
Peak Amplitude –6 dBm
Duty Cycle 25%