Figure 2- 4 shows that how you can apply a Channel Offset or a 
Frequency Dependent Offset prior to any mathematical functions. These 
allow you to compensate each channel individually. An overall offset can 
be applied if required using the Display Offset.
Setting Channel Offsets
This gain or loss is applied to the measured power before any 
mathematical functions, display offsets or relative functions are included. 
Offsets are entered in dB and the range is –100 dB to +100 dB. 
The equivalent range in % is from 0.0000000001% to 10000000000%.
A positive value compensates for a loss and a negative value compensates 
for a gain. 
Procedure 
To enter a channel offset:
1 Press   to display the Channel Setup screen. Confirm the channel 
requiring setup is displayed.
2 Press   to display the Offsets Setup.
3 Use the   and   keys to highlight the Offset setting field.
4 Press   to check the Offset setting field.
Figure 2-5 Typical channel offset display
5 Press   to highlight the Offset value field and press   to display 
the  Offset pop- up. Use the numeric keypad to enter the required value 
in the Offset pop- up window.