User`s guide
General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 41
Setting Frequency Dependent Offsets
Frequency dependent offset tables provide a quick and convenient method 
of compensating for frequency related changes in the response of your test 
system. Note that when selected, frequency dependent offset corrections 
are applied IN ADDITION to any correction for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset 
tables with a maximum of 80 frequency points each.
To use frequency dependent offset tables:
1 Select the table to be applied to a channel. Refer to “Setting Frequency 
Dependent Offsets” on page 41 for further information. If you require to 
edit the table refer to “Editing Frequency Dependent Offset Tables” on 
page 44 for further information.
2 If an 8480 Series, N8480 Series, an E- Series, or an U2000 Series sensor 
is used, zero and calibrate the power sensor. The reference calibration 
factor used during the calibration is automatically set by the power 
meter from the sensor calibration table (if selected).
3 Specify the frequency of the signal you want to measure. The 
calibration factor/offset is automatically set by the power meter from 
the sensor calibration table (if selected) and the frequency dependent 
offset table. Refer to “Procedure” on page 42 for further information.
4 Make the measurement.
Selecting a Frequency Dependent Offset Table
You can select a frequency dependent offset table from the   key 
menu followed by  ,  , and   for Channel A or 
Channel B.
For Channel C, you can select the frequency dependent offset table from 
the   key menu followed by  ,  , and  .
For Channel D, you can select the frequency dependent offset table from 
the   key menu followed by  ,  , and  .
Tab les
Meter
Freq. Dep. Offset
Tab les
Sensor ChC
Freq. Dep. Offset
Tab les
Sensor ChD
Freq. Dep. Offset










