Procedure
Set duty cycle as follows:
1 Press   to display the Channel Setup screen. Confirm the channel 
requiring setup is displayed.
2 Press   to display the Offsets Setup.
3 Use the   and   keys to highlight the Duty Cycle setting field.
4 Press   to check the Duty Cycle setting field.
Figure 2-17 Duty cycle setting
5 Press   to highlight the Duty Cycle value field and press   to 
display the Duty Cycle pop- up. Use the numeric keypad to enter the 
required value in the Duty Cycle pop- up window.
6 Confirm your choice by pressing  .
7 Press   key to complete the duty cycle entry. The Dty Cyc 
indicator is displayed as shown in Figure 2- 18.
Figure 2-18 Duty cycle indicator
Offsets
Duty cycle 
setting field
Duty cycle 
value field
%
Duty cycle 
indicator
NOTE
Pulse power averages out any aberrations in the pulse such as overshooting or ringing. 
For this reason, it is called pulse power and not peak power or peak pulse power.
In order to ensure accurate pulse power readings, the input signal must be pulsed with a 
rectangular pulse. Other pulse shapes (such as triangle, chirp, or Gaussian) will cause 
erroneous results.