User`s guide
2 General Power Meter Functions
58 N1913/1914A EPM Series Power Meters User’s Guide
12 Press   or   to setup measurement setting such as 
measurement averaging, measurement frequency, offsets, duty cycle and 
so forth. Refer to “General Power Meter Functions” on page 31 for the 
setup procedures.
13 Send  *OPC (OPeration Complete) command to meter to set the 
operation complete bit in the Standard Event Status register when all 
pending device operations are completed.
14 Set the meter to continuous trigger mode by sending 
INITiate:CONTinuous ON command to meter.
15 Configure and set the required power sweep range and step on the 
power source accordingly. 
16 Set the power source trigger input and trigger output, then start 
sweeping.
17 Poll the status of the power meter by sending *ESR?. *ESR? will return 
a 1 when buffering is completed. Use FETCh? to retrieve all the 
buffered measurement.
NOTE
Refer to the N1913/1914A EPM Series Power Meters Programming Guide for the 
commands usage details.
Example of command set: 
BUFF:COUN 100
NOTE
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488) 
programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation 
Guide for details.