User`s guide
2 General Power Meter Functions
60 N1913/1914A EPM Series Power Meters User’s Guide
10 Press   or   to setup measurement settings such as 
measurement averaging, measurement frequency, offsets, duty cycle and 
so forth. Refer to “General Power Meter Functions” on page 31 for the 
setup procedures.
11 Set the frequency range and step by sending the below commands to 
meter using remote interface. 
• SENSe1:FREQuency:STARt 
<start_frequency><frequency_unit>
• SENSe1:FREQuency:STOP <stop_frequency><frequency_unit>
• SENSe1:FREQuency:STEP <frequency_step_size>
12 Send  *OPC (OPeration Complete) command to meter; to set the 
operation complete bit in the Standard Event Status register when all 
pending device operations are completed.
13 Set the meter to continuous trigger cycles by sending 
INITiate:CONTinuous ON command to meter.
14 Configure and set the required power sweep range and step of the 
power source. 
15 Set the power source trigger input and trigger output. 
16 Set the power source to sweep.
NOTE
Refer to the EPM Series Power Meters Programming Guide for the commands usage 
details.
Example of commands set:
FREQ:STAR 10MHz
FREQ:STOP 500MHz
FREQ:STEP 10
NOTE
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488) 
programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation 
Guide for details.