Agilent Technologies 53150A / 53151A / 53152A
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
7
Änderungen und Irrtümer vorbehalten. dataTec 05-03-2014 Update Agilent: December 5, 2011 | 5966-3118E
7
Timebase
Frequency:10MHz
Output:10MHzsinewave,1Vrmsinto
50Ω
Externaltimebaseinput:1,2,5,10MHz;
1to5Vrmsinto50Ω
Connector:BNCfemalelocatedonrear
panel
Internaltimebasestability
TCXO
(Standard)
Oven
(Option001)
Agingrate
Perday — <5x10
–10
Permonth <1x10
–7
<1.5x10
–8
Shortterm<1x10
–9
<2x10
–10
(1savg.time)
Line variation<1x10
–7
<1x10
–10
(±10%)
Warm-up — <1x10
–8
within5min.
after turn-on
at25°C
Temperature
stability(0-55°C) <1x10
–6
<1x10
–8
Generalinformation
Saveandrecall:Upto9complete
instrumentsetupsmaybesavedand
laterrecalled.Thesesetupsare
retainedwhenpowerisremoved.
Samplerate:User-selectable
Fast(nominally20msbetween
readings),Medium(nominally
250msbetweenreadings),Slow
(nominally1sbetweenreadings)
andHold.
Selftest:Internalmemoryandcount
circuitry automatically tested at
startup,viamenuselection,or
remotely.Errormessagesdisplayed
toindicatefailedtests.
Size:213mmWx88.5mmHx300mmD
Operatingtemperature:0-55°C
Withbatteryoption:0-40°C
Weight:4kgwithoutbatteryoption,
6.4kgwithbatteryoption
Warranty: 1year
Programming:GPIB(IEEE-488.1-1987,
IEEE488.2-1987)orRS-232C
Language:SCPI-1992.0(Standard
CommandsforProgrammable
Instruments)
RS-232Crates:User-selectable2400
to19200baud
Powersupply:
AC:90-132Vac;47.5-66Hzor360-
440Hz216-264Vac;47.5-66Hz
line selection: automatic
powerrequirements:75VAmax.
(25Wtyp.)
DC:(Option002only):11-18Vdc;
2Amax.
Battery(Option002):
Type:VHScamcorder,leadacid
(2each)
Chargetime:8hoursinunit
Capacity:2.5hoursmin.at25°C
Math functions:
Offset: Lastreadingand/or
enteredoffsettoreadingforeither
powerorfrequency
Averaging:1to99measurement
runningaverage
Cablelosscompensation: Offsets
powerreadingvialinear
interpolationofuser-entered
attenuationswithupto9
independentfrequencypoints.
Display:BacklitLCD.Backlightcanbe
turnedonoroffviafrontpanelcon-
trol.
Sleepmode(Option002only):
Automaticallyactivatedifnoinputis
presentfor5minutes.
Safety:Designedincompliancewith
IEC-1010,CAN/CSA1010.1
EMC:Designedincompliancewith
IEC-11,EN50082-1,IEC801-2,-3,-4.
Accessoriessupplied
Operating,programming,andservice
manuals and ACpowercord.
Orderinginformation
Agilent53150A 20GHzcounter
Agilent53151A 26.5GHzcounter
Agilent53152A 46GHzcounter
Options
Option001 OvenTimebase
Option002 BatteryandDCinput
Option1CM
(53150-67001)Rackmountkit
53153A(Option007)Softcarryingcase
for
5315xAmicrowavecounters
OptionA6J ANSI7540Compliant
Calibration
OptionW30 ThreeyearsofReturnRepair
Service
OptionW50 FiveyearsofReturnRepair
Service