Specifications
Waveform Math hthis exercisetheWAVEFORM MATH menuL usedto subtra∝
one
Exercise WavefOmfrom another.
:nstrument Setup setup an HP 8116へ
or a signalgenerator capabL ofa l ML l v01t
squarewave,as fonows:
o MODE=NORM
o FRQ=l MHz
●
 DTY=50%
o AMP=1.00V
o OFS=0.00V
e Set the generator for a square wave slgnal
Comcd thcsignalto a BNC tec on channell usmg a l‐
metcr coai」
cableo Connect anothcr l‐meter cable frOm thc other sidc Ofthe BNC tee
and terminate in 50 Ω to channe1 2.
Osc:::oscope Setup Thc cxtra cablelem betwecn chalmels l and 2 prondes atime delay
between the stgnals on the osttoscOpe. The propagation Of a l_meter
coぶ
al cable is appropamatdy 6 to 7 ns.This dclay is uscd to demonstrate
the math ttction.
The following 
procedure 
assists 
in 
setting 
up 
the 
HP 
545104 for 
optimal
viewing.
o  Press 
ALITOSCALE.
o  Press DISPLAY to set the best 
viewing 
conditions. 
Set 
display mode
to avg, # of avgto 8, 
# 
of screens 
to 2.
o 
Press WFORM 
MATH 
to define 
the function. 
Select/  and 
turn the
display on. Select 
chan 1, 
- 
(minus), 
chan 2 
and set the function
sensitivity to 
2.00 
V/div.
HP 54510A
Front‐
Panel Reference
Waveform Math Menu
9‐7