Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
Tektronix TDS 3012B Oscilloscope 
FFT Math Function 
  This function converts an oscilloscope time-domain waveform into its frequency 
spectrum. Waveforms with a D.C. component may cause errors in FFT magnitudes. 
  Push the Save/Recall button, Recall Factory Setup button and the Confirm button. 
Connect the source to the scope and display the time-domain waveform. The acquisition mode 
should be set to sample and the horizontal resolution should be set to normal. These are found 
in the Acquire menu. The Acquire menu also displays the sample rate. Push the CH 1 and 
menu buttons. AC coupling is preferred. The bandwidth can be set to 20 MHz. This filters the 
source signal, limiting frequencies below the nyquist frequency. Full band width is 150 MHz. 
Use the fine scale button so the waveform uses approx. 95% of the vertical scale. The horizontal 
scale knob should be set to display a minimum of five periods. This controls the sampling 
frequency and the frequency resolution. The more cycles displayed increases the frequency 
resolution and decreases the sampling rate. This is the trade off. The nyquist frequency is half the 
sampling frequency. Aliasing may occur if the sampling rate is low. Push the math and FFT 
buttons for the frequency spectrum. The rectangular window is the only non-windowing 
function. Non-windowing is used for transients. Periodic waveforms require windowing. The 
magnification button is used along with the horizontal scale and position knobs. This is used to 
magnify and position the FFT spectrum. Magnification is about the center of the display. The 
entire waveform record is used to calculate the FFT spectrum. Magnification increases display 
resolution only. The FFT spectrum is not recalculated. Recall the factory setup before a new FFT 
spectrum is calculated. 
Examples: 
  1GS/s = sample rate for display 
  Nyquist = 1GHz / 2 = 500 MHz BW 
  500 MHz / 10 div = 50 MHz / div 
  freq. resolution = 50 MHz / 50 = 1 MHz. 
  50 MS/s = sample rate for display. 
  Nyquist = 50 MHz / 2 = 25 MHz BW 
  25 MHz / 10 div = 2.5 MHz / div 
  freq. resolution = 2.5 MHz / 50 = 50 kHz 










