Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
Main / Delayed menu 
  Main is the normal viewing mode for the display. Delayed mode expands the main view. 
In Delayed mode the top half of the display shows the main sweep and the bottom half shows the 
delayed sweep. The delayed sweep is a magnified portion of the main sweep. The magnified 
portion is adjustable using the Horizontal controls. 
Single and Run Control 
  The continuous Run mode uses half the memory to acquire data and the other half is 
used to process the previous acquisition. 
  Single acquisition mode uses twice the memory as the Run acquisition mode. The 
increase in memory allows for a high sampling rate at slower sweep speeds. Single acquisition 
will acquire data with the longest record. So use the Single acquisition mode for maximum 
sampling rate and memory depth. Set the trigger mode to Normal and the trigger level high 
enough so the desired event will trigger the acquisition and not a low level event. The scope will 
otherwise immediately trigger. 
Acquire Modes 
Normal > Best for most waveforms. No averaging. Default setting. 
 > At slow speeds extra samples are discarded. 
Peak Detect > Best for capturing narrow pulses. 
> At slow speeds extra samples are kept to capture narrow pulses. 
> Displays pulses as narrow as 500 pS. 
High Resolution > At slow speeds extra samples are averaged to produce a point. 
> Limits the scopes real – time bandwidth (acts as a low – pass filter). 
Average > Averages acquisitions to reduces noise. 
 > Bandwidth and rise time are not reduced. 